Calipso

Sirolimus Eluting Сoronary Stent System
Calipso
Calypso coronary stent with bioresorbable drug coating on the delivery system

Coronary stent with a completely bioresorbable coating containing sirolimus (rapamycin) - a highly effective drug agent for preventing restenosis.

The stent system consists of a delivery system equipped with a balloon (expanding element) located at the distal end of the system, and a stent with a bioresorbable drug coating installed and fixed on the delivery system over the balloon. The drug coating of the stent is a mixture of the drug sirolimus and a copolymer of glycolic and lactic acids.
Intended use
The stent system is designed to fix the lumen of a blood vessel during coronary angioplasty (treatment of stenosis in the coronary blood vessels of the heart) by expanding the lumen of the vessel at the site of stenosis.
Stent characteristics
Material
Cobalt Chrome L605
Construction
Matrix tubular
Coating
Sirolimus (rapamycin)
Delivery system characteristics
Delivery system type
Rapid Exchange (Rx)
Compatibility with coronary guidewire
0.014”
Compatibility with guiding catheter
5F
Balloon material
Pebax
Nominal pressure
9 atm
Rated burst pressure
18 atm. for stents ⌀ 2.0-4.5 mm.
Size range
Array

  8 11 12 13 15 18 22 23 28 33 38
2.0 KL08200 KL11200   KL13200 KL15200 KL18200   KL23200 KL28200    
2.25 KL08225 KL11225   KL13225 KL15225 KL18225   KL23225 KL28225 KL33225 KL38225
2.5 KL08250 KL11250 KL12250 KL13250 KL15250 KL18250 KL22250 KL23250 KL28250 KL33250 KL38250
2.75 KL08275 KL11275 KL12275 KL13275 KL15275 KL18275 KL22275 KL23275 KL28275 KL33275 KL38275
3.0 KL08300 KL11300 KL12300 KL13300 KL15300 KL18300 KL22300 KL23300 KL28300 KL33300 KL38300
3.5 KL08350 KL11350 KL12350 KL13350 KL15350 KL18350 KL22350 KL23350 KL28350 KL33350 KL38350
4.0 KL08400 KL11400 KL12400 KL13400 KL15400 KL18400 KL22400 KL23400 KL28400 KL33400 KL38400
4.5 KL08450 KL11450   KL13450 KL15450 KL18450 KL22450 KL23450 KL28450 KL33450 KL38450